Junzi Cao1, Wen Sun, Feifei Gong
1School of Pharmacy, Yantai University, Yantai, P. R. China.
Abstract:
CIEF was developed for the rapid analysis of charge heterogeneity of trastuzumab biosimilar using commercially available fluorocarbon-coated capillary. The CIEF master mix was composed of 0.30% w/v methyl cellulose, 2.3 M urea, 56.8 mM l-arginine, 1.52 mM iminodiacetic acid, 4.5% v/v carrier ampholytes (broad-range pI 3-10 and narrow-range pI 8-10.5 with ratio of 3:1), and 0.45% v/v 10.0, 9.5, 7.0, 5.5, 4.1 pI markers. To get a robust method to analyze charge heterogeneity, some separation parameters, including focusing time and separation temperature, were investigated and optimized. The optimized method gave good precision in estimated pI values of charge variants with RSDs of not more than 0.16% intraday analysis (n = 6) and < 0.18% interday analysis (n = 9). In addition, the applications of this method including purity, stability, lot consistency, peptide N-glycosidase F digest, and C-terminal lysine variants characterization were also investigated.