Related Experiment Video For Colon
Updated: May 2, 2026

Murine Endoscopy for In Vivo Multimodal Imaging of Carcinogenesis and Assessment of Intestinal Wound Healing and Inflammation
Published on: August 26, 2014
Forward-viewing radial-array echoendoscope for staging of colon cancer beyond the rectum
Pradermchai Kongkam1, Sittikorn Linlawan1, Satimai Aniwan1
1Pradermchai Kongkam, Sittikorn Linlawan, Satimai Aniwan, Narisorn Lakananurak, Suparat Khemnark, Wiriyaporn Ridtitid, Pinit Kullavanijaya, Rungsun Rerknimitr, Gastrointestinal Endoscopy Excellent Center, Department of Medicine, King Chulalongkorn Memorial Hospital, Chulalongkorn University, Bangkok 10330, Thailand.
Aim:
To evaluate feasibility of the novel forward-viewing radial-array echoendoscope for staging of colon cancer beyond rectum as the first series.
Methods:
A retrospective study with prospectively entered database. From March 2012 to February 2013, a total of 21 patients (11 men) (mean age 64.2 years) with colon cancer beyond the rectum were recruited. The novel forward-viewing radial-array echoendoscope was used for ultrasonographic staging of colon cancer beyond rectum. Ultrasonographic T and N staging were recorded when surgical pathology was used as a gold standard.
Results:
The mean time to reach the lesion and the mean time to complete the procedure were 3.5 and 7.1 min, respectively. The echoendoscope passed through the lesions in 13 patients (61.9%) and reached the cecum in 10 of 13 patients (76.9%). No adverse events were found. The lesions were located in the cecum (n = 2), ascending colon (n = 1), transverse colon (n = 2), descending colon (n = 2), and sigmoid colon (n = 14). The accuracy rate for T1 (n = 3), T2 (n = 4), T3 (n = 13) and T4 (n = 1) were 100%, 60.0%, 84.6% and 100%, respectively. The overall accuracy rates for the T and N staging of colon cancer were 81.0% and 52.4%, respectively. The accuracy rates among traversable lesions (n = 13) and obstructive lesions (n = 8) were 61.5% and 100%, respectively. Endoscopic ultrasound and computed tomography had overall accuracy rates of 81.0% and 68.4%, respectively.
Conclusion:
The echoendoscope is a feasible staging tool for colon cancer beyond rectum. However, accuracy of the echoendoscope needs to be verified by larger systematic studies.
More Related Videos
03:43Structured Approach to Colonoscopy Technique Optimization: A Single-Center Experience with Novice Endoscopists
Published on: July 11, 2025
15:49Flexible Colonoscopy in Mice to Evaluate the Severity of Colitis and Colorectal Tumors Using a Validated Endoscopic Scoring System
Published on: October 16, 2013
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Endoscopic Procedures II: Colonoscopy
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures III: Video Capsule Endoscopy