Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Demonstration of compatibility of multiple arterial blood gas syringes with current multi-parameter analyzers
Laura E Seiberlich1, Lisa M Cifaldi
1Smiths Medical ASD Inc. , St Paul, MN , USA.
Objective:
Since there is a lack of current evidence to support the compatibility of the most commonly used arterial blood gas syringes with the latest multi-parameter analyzers, the objective of this study was to assess the agreement of analyte values between three heparinized arterial blood gas syringes using three different analyzers.
Methods:
Venous blood from 25 healthy volunteers was drawn into the study syringes (BD Drihep A-Line, PICO50, and Portex Line Draw Plus) by four clinician volunteers in a random order (441 total) and immediately (<15 minutes) analyzed (ABL800 FLEX, RAPIDPoint 500, i-STAT System) in a randomized order. The Bland-Altman method was used to assess agreement of analyte values between the syringes for each analyzer. The results by analyte and analyzer were compared across syringes using one-way ANOVA, and Tukey's approach was used to identify statistically significant differences between pairs of syringes.
Results:
Comparison of the syringes' mean differences and standard deviations showed close agreement for all three analyzers. There were no statistically significant differences between syringes in 14 of the analytes for any of the analyzers (p > 0.05). For RAPIDPoint and i-STAT, the Ca(2+) value was significantly less for A-Line than for the other syringes. The value for Na(+) was significantly greater for the PICO50 than the A-Line and Line Draw syringes with the i-STAT. Both results were within two standard deviations of the mean of the other two syringes and are not considered clinically significant; however, comparisons were not made between the values from the different analyzers.
Conclusions:
Dry-balanced lithium heparin ABG syringes used for blood draws should provide reliable results, regardless of syringe type, provided that the clinicians use proper pre-analytical techniques.
More Related Videos
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
Related Concept Videos
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...