Load-transfer analysis after insertion of cementless anatomical femoral stem using pre- and post-operative CT images based patient-specific finite element analysis
Related Experiment Video
Updated: May 2, 2026

Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
Go Yamako1, Etsuo Chosa2, Xin Zhao3
1Organization for Promotion of Tenure Track, University of Miyazaki, Miyazaki, Japan.
Abstract:
Periprosthetic bone remodeling is commonly seen after total hip arthroplasty, but the remodeling pattern differs among patients even in those implanted with the same stem. Remodeling occurs mainly because of the difference in load transmitted from the stem to the femur. In this study, we evaluated the load-transfer pattern in eight female patients implanted with an anatomical stem on an individual basis by patient-specific finite element analysis that is based on pre- and postoperative computed tomography images. Load transfer was evaluated using interface stress between the stem and bone. One of eight patients demonstrated proximal dominant load transfer, while the other patients demonstrated a distal dominant pattern. The results of our biomechanical simulations reveal the differences in load-transfer pattern after surgery among patients with the same anatomical stem.