Effect of localizer radiograph on radiation dose associated with automatic exposure control: human cadaver and
Sarabjeet Singh1, Dean Petrovic, Ethen Jamnik
1From the *Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Boston, MA; and †Department of Radiology, Ljubljana University Medical Center, Ljubljana, Slovenia.
Purpose:
To evaluate the effect of localizing radiograph on computed tomography (CT) radiation dose associated with automatic exposure control with a human cadaver and patient study.
Materials And Methods:
Institutional review board approved the study with a waiver of informed consent. Two chest CT image series with fixed tube current and combined longitudinal-angular automatic exposure control (AEC) were acquired in a human cadaver (64-year-old man) after each of the 8 combinations of localizer radiographs (anteroposterior [AP], AP lateral, AP-posteroanterior [PA], lateral AP, lateral PA, PA, PA-AP, and PA lateral). Applied effective milliampere second, volume CT dose index (CTDIvol) and image noise were recorded for all 24-image series. Volume CT dose indexes were also recorded in 20 patients undergoing chest and abdominal CT after PA and PA-lateral radiographs with the use of AEC. Data were analyzed using analysis of variance and linear correlation tests.
Results:
With AEC, the CTDIvol fluctuates with the number and projection of localizer radiographs (P < 0.0001). Lowest CTDIvol values are seen when 2 orthogonal localizer radiographs are acquired, whereas highest values are seen when single PA or AP-PA projection localizer radiographs are acquired for planning (P < 0.0001). In 20 patients, CT scanning with AEC after acquisition of 2 orthogonal projection localizer radiographs was associated with significant reduction in radiation dose compared to PA projection radiographs alone (P < 0.0001).
Conclusions:
When scanning with AEC, acquisition of 2 orthogonal localizer radiographs is associated with lower CTDIvol compared to a single localizer radiograph.