Reliability and validity of isometric knee extensor strength test with hand-held dynamometer depending on its
Won Kuel Kim1, Don-Kyu Kim1, Kyung Mook Seo1
1Department of Physical Medicine and Rehabilitation, Chung-Ang University College of Medicine, Seoul, Korea.
Annals of Rehabilitation Medicine
|March 19, 2014
Summary
Fixing a hand-held dynamometer (HHD) during isometric knee extensor strength tests significantly improves its reliability and validity compared to non-fixed methods. This validated approach enhances quadriceps strength measurement accuracy.
Area of Science:
- Biomechanics
- Musculoskeletal assessment
- Rehabilitation technology
Background:
- Accurate measurement of isometric knee extensor strength is crucial for assessing muscle function and guiding rehabilitation.
- Hand-held dynamometers (HHDs) offer a portable alternative to isokinetic dynamometers, but their measurement reliability and validity are debated.
- The impact of device fixation on HHD performance in strength testing requires clarification.
Purpose of the Study:
- To evaluate the reliability and validity of a fixed versus non-fixed hand-held dynamometer (HHD) for measuring isometric knee extensor strength.
- To compare HHD measurements against a gold-standard isokinetic dynamometer.
Main Methods:
- Twenty-seven healthy females performed isometric knee extensions using an isokinetic dynamometer, a non-fixed HHD, and a fixed HHD (secured with Velcro).
- Measurements were taken in both seated and supine positions with the knee fixed at 35°.
- Peak torque values from three repetitions were analyzed for reliability and validity.
Main Results:
- The fixed HHD demonstrated significantly higher validity and reliability compared to the non-fixed HHD when correlated with isokinetic dynamometer results.
- Pearson correlation coefficients were significant for the fixed HHD (r=0.473-0.806) but largely non-significant for the non-fixed HHD.
- Both methods showed excellent inter-rater reliability, with the fixed HHD exhibiting slightly higher intraclass correlation coefficients (0.952-0.984 vs. 0.940-0.963).
Conclusions:
- Fixating the hand-held dynamometer during isometric knee extensor strength testing, particularly in the supine position, enhances measurement reliability and validity.
- This finding supports the use of a fixed HHD for more accurate quadriceps strength assessments in clinical and research settings.
Keywords:
Dimensional measurement accuracyIsometric contractionKnee jointMuscle strength dynamometerPhysical restraint