Systematic evaluation of commercial susceptibility testing methods for determining the in vitro activity of
Thomas J Kirn1, Elizabeth Onyeaso2, Madiha Syed2
1Department of Pathology and Laboratory Medicine, Rutgers Robert Wood Johnson Medical School, New Brunswick, New Jersey, USA Department of Medicine (Infectious Disease), Rutgers Robert Wood Johnson Medical School, New Brunswick, New Jersey, USA kirntj@rutgers.edu.
Abstract:
We systematically evaluated 5 methods for testing daptomycin versus 48 Enterococcus faecalis, 51 Enterococcus faecium, and 50 Staphylococcus aureus isolates using (i and ii) broth microdilution (BMD) with 50-mg/liter calcium medium supplementation (reference method) and 30-mg/liter calcium medium supplementation (BMD30 method), (iii) Etest, and (iv and v) MicroScan panel 33 using 2 methods to prepare the bacterial inoculum (MicroScan turbidity and MicroScan Prompt). Isolates were categorized as susceptible (S) or nonsusceptible (NS) based on measured MICs. Essential (± 1 dilution) agreement (EA) and categorical (S/NS) agreement (CA) for each method were compared to the reference method. For E. faecium, categorical agreement was poor between the reference method and BMD30 as well as with the three commercial methods, with frequent false-NS results (30 for BMD30, 18 for Etest, 22 for MicroScan Prompt, and 25 for MicroScan turbidity). All E. faecalis isolates were judged to be S by the reference method; two of these isolates were categorized as NS using the BMD30 method, and one was categorized as NS by all three commercial methods. All S. aureus isolates were judged to be S using all five methods. MIC values determined by the comparator methods tended to be higher than those for the reference method, especially for E. faecium isolates. EAs between the reference BMD and BMD30, Etest, MicroScan Prompt, and MicroScan turbidity were 63%, 63%, 63%, and 56%, respectively, for E. faecium, 87%, 83%, 98%, and 80%, respectively, for E. faecalis, and all 100% for S. aureus.
More Related Videos
11:28A Reference Broth Microdilution Method for Dalbavancin In Vitro Susceptibility Testing of Bacteria that Grow Aerobically
Published on: September 9, 2015
09:57System for Efficacy and Cytotoxicity Screening of Inhibitors Targeting Intracellular Mycobacterium tuberculosis
Published on: April 5, 2017
Related Concept Videos
Antimicrobial Effectiveness
Automated Microbial Diagnostics
Clinical Significance of Antibiotic Resistance