Evaluation of Left Ventricular Structure and Function using 3D Echocardiography
Published on: October 28, 2020
Intraoperative three-dimensional versus two-dimensional echocardiography for left ventricular assessment
Alessandra Meris1, Luisa Santambrogio, Gabriele Casso
1From the Division of Cardiothoracic Anesthesiology, Fondazione Cardiocentro Ticino, Lugano, Switzerland.
Three-dimensional transesophageal echocardiography (3D TEE) offers comparable left ventricular (LV) function assessment to 2D TEE in cardiac surgery. While 3D TEE yields larger LV volumes, it does not alter clinical classification, ensuring similar diagnostic accuracy.
Area of Science:
- Cardiology
- Medical Imaging
- Echocardiography
Background:
- Three-dimensional transesophageal echocardiography (3D TEE) is increasingly utilized in intraoperative cardiac surgery.
- 3D TEE demonstrates better correlation with cardiac magnetic resonance imaging for left ventricular (LV) measurements than 2D transthoracic echocardiography.
Purpose of the Study:
- To compare intraoperative 3D TEE with conventional 2D TEE for quantitative assessment of LV function.
Main Methods:
- 156 patients undergoing elective cardiac surgery had both 2D TEE and 3D TEE performed.
- LV volumes and ejection fraction (EF) were calculated using standard methods for both techniques.
- Quantitative indices, image quality, acquisition/analysis times, and reproducibility were compared using statistical tests and Bland-Altman analysis.
Main Results:
- No significant difference was found in LV ejection fraction (EF) estimation between 3D and 2D TEE (P = 0.227).
- 3D TEE calculated larger indexed LV end-diastolic and end-systolic volumes (P < 0.001), but this did not alter clinical classification in most cases (98.8% for iEDV, 92.8% for iESV).
- Image quality and acquisition time were similar, but 3D TEE analysis required significantly more time (P < 0.001). Reproducibility was comparable for both methods.
Conclusions:
- Intraoperative 3D TEE provides comparable LV global function quantification, acquisition time, and reproducibility to 2D TEE.
- Despite calculating larger LV volumes, 3D TEE measurements do not affect the clinical classification of LV dilation.
- 3D TEE is a viable alternative for intraoperative LV assessment, though analysis time is longer.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Cardiac Catheterization III: Left Heart Catheterization