Point-of-care fluorescence immunoassay for cardiac panel biomarkers
Tae Kyum Kim1, Sang Wook Oh, Soon Cheol Hong
1Department of Biomedical Science, Hallym University, Chuncheon, South Korea; BodiTech Med, Inc, Chuncheon, South Korea.
Background:
Myoglobin, creatine kinase-MB isoenzyme (CK-MB), and cardiac troponin I (cTnI) are cardiac biomarkers that are widely used to assist in the early and late diagnoses of acute myocardial infarction (AMI). Here, we present a clinically applicable fluorescence (FL) immunoassay for cardiac biomarkers.
Methods:
Whole blood was mixed with FL-labeled detector Ab (antibody) and then loaded onto a capture Ab-immobilized strip in a test cartridge. The FL intensities at test and control line on the strip were obtained and converted in a laser FL scanner to determine the concentration of biomarker. The analytical performance of immunoassay system was evaluated by linearity and imprecision tests. The comparability of the FL immunoassay method was examined with a reference method.
Results:
FL intensities and the levels of myoglobin, CK-MB, and cTnI displayed good linearity and high correlations (r = 0.999, 0.998, and 0.989, respectively). The coefficient of variations (CVs) of imprecision for all cardiac biomarkers were less than 8% in both intra- and interassays. When the results from the developed method and bioMerieux VIDAS assay were analyzed by Bland-Altman plot and Passing-Bablok plot, the two assay methods were in good agreement.
Conclusion:
The FL immunoassay system can provide a platform for point-of-care testing (POCT), and it is an easy, fast, and reliable method for the quantification of cardiac biomarkers.
Related Concept Videos
Immunofluorescence Microscopy
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...