Related Experiment Video
Updated: May 1, 2026

Nanosensors to Detect Protease Activity In Vivo for Noninvasive Diagnostics
Published on: July 16, 2018
Improving detection specificity of iron oxide nanoparticles (IONPs) using the SWIFT sequence with long T(2)
Luning Wang1, Wei Tang2, Zipeng Zhen2
1Department of Physics and Astronomy, University of Georgia, Athens, US; BioImaging Research Center (BIRC), University of Georgia, Athens, US.
Abstract:
In order to improve the detection specificity of iron oxide nanoparticles (IONPs) delivered to tumors, we embedded saturation pulses into the sweep imaging using Fourier transformation (SWIFT) sequence to suppress long T(2) tissues and fat. Simulation of the Bloch equation was first conducted to study behavior of the saturation pulses of various lengths under different T(2) and off-resonance conditions. MR experiments were then conducted using in vivo mouse xenografts and a phantom consisting of IONPs, vegetable oil, and explanted tumor specimen, without and with long T(2) suppression under a 7T magnetic field. For the in vivo study, arginine-glycine-aspartate (RGD) coated 10nm IONPs (RGD-IONPs) were delivered to tumors implanted in nude mice through both intra-tumor and intravenous injections. Histological studies confirmed that RGD-IONPs efficiently homed to tumors through RGD-integrin interaction. Compared to conventional SWIFT, the proposed method resulted in sufficient suppression on long T(2) species but less influence on short T(2) species. For both the in vivo and ex vivo studies, significantly improved contrast-to-noise ratio (CNR) was achieved between the IONPs and the long T(2) species.