Related Experiment Video
Updated: May 1, 2026

Analyzing Long-Term Electrocardiography Recordings to Detect Arrhythmias in Mice
Published on: May 23, 2021
Trying to predict the unpredictable: Variations in device-based daily monitored diagnostic parameters can predict
Ewa Jędrzejczyk-Patej1, Oskar Kowalski, Beata Sredniawa
1Department of Cardiology, Congenital Hear t Disease and Electr otherapy, Silesian Medical Univer sity, Silesian Centre for Hear t Diseases, Zabrze, Poland. ewajczyk@op.pl.
Background:
The aim of this study was to evaluate the value of device-based diagnostic parameters in predicting ventricular arrhythmias in cardiac resynchronization therapy (CRT) recipients.
Methods:
Ninety-six CRT-D patients participating in TRUST CRT Trial were analyzed. The inclusion criteria were: heart failure in NYHA ≥ 3 class, QRS ≥ 120 ms, LVEF £ 35% and significant mechanical dyssynchrony. Patients were divided into those with (n = 31, 92 arrhythmias) and without (n = 65) appropriate ICD interventions within follow-up of 12.03 ± 6.7 months. Daily monitored device-based parameters: heart rate (HR), thoracic impedance (TI), HR variability and physical activity were analyzed in 4 time windows: within 10, 7, 3 days and 1 day before appropriate ICD interventions.
Results:
A consistent pattern of changes in three monitored factors was observed prior to arrhythmia: 1) a gradual increase of day HR (from 103.43% of reference within 10-day window to 105.55% one day before, all p < 0.05 vs. reference); 2) variations in night HR (104.75% in 3 days, 107.65% one day before, all p < 0.05) and 3) TI decrease (from 97.8% in 10 days to 96.81% one day before, all p < 0.05). The combination of three parameters had better predictive value, which improved further after exclusion of patients with atrial fibrillation (AF). The predictive model combining HR and TI together with LVEF and NT-proBNP was more prognostic than the model involving LVEF and NT-proBNP alone (difference in AUC 0.05, 95% CI 0.0005-0.09, p = 0.04).
Conclusions:
Daily device-monitored parameters show significant variations prior to ventricular arrhythmia. Combination of multiple parameters improves arrhythmia predictive performance by its additive value to baseline risk factors, while presence of AF diminishes it.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias V: Evaluating Dysrhythmias
Mechanism of Cardiac Arrhythmias
Holter Monitor: 24-Hour Monitoring
Dysrhythmias VII: Nursing Management of Dysrhythmias
Dysrhythmias VI: Management of Dysrhythmias