Linear Measurement Accuracy of Eight Cone Beam Computed Tomography Scanners
Pasupen Kosalagood1, Onanong C Silkosessak1, Pisha Pittayapat1,2
1Department of Radiology, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand.
Background:
Information regarding linear accuracy is necessary for efficient treatment evaluation, especially for maxillofacial reconstruction or implants.
Purpose:
To investigate the accuracy of linear measurements from multiple cone beam computed tomography (CBCT) devices.
Materials And Methods:
A RANDO® phantom was scanned with eight CBCT scanners (11 modes). The viewing software accompanying each scanner was employed for measurements in mediolateral, anteroposterior, and supero-inferior dimensions by two dentomaxillofacial radiologists. Digital caliper measurements were used as a "gold standard." ANOVA with Scheffé post hoc analysis and intraclass correlation coefficient (ICC) were utilized for statistical analyses. The level of confidence was 95%.
Results:
Differences from the gold standard among 11 acquisition modes were statistically significant (p < .001). Measurements from one unit were always underestimated compared with all others (p < .001). The range of absolute measurement errors for tested units was -2.56 to 0.54 mm (mean ± SD 0.45 ± 0.71) including the outlier and -0.34 to 0.54 mm (0.16 ± 0.11) excluding the outlier. Slightly more values were underestimated than overestimated (41 of 66 measurements, 7 out of 11 CBCT modes). ICC scores for inter- and intraobserver agreement were perfect (1.000).
Conclusions:
Treatment planning from large-volume CBCT was found to be reliable in all except one of the investigated scanners. New CBCT scanners should always be tested for geometric accuracy.
More Related Videos
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Adjusting a Traverse
Accuracy and Precision
Imaging Studies III: Computed Tomography