Radiation dose and contralateral breast cancer risk associated with megavoltage cone-beam computed tomographic image
Alexandra Quinn1, Lois Holloway2, Eng-Siew Koh3
1Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia; Liverpool and Macarthur Cancer Therapy Centres, NSW, Australia.
Purpose:
To measure and compare organ doses from a standard tangential breast radiation therapy treatment (50 Gy delivered in 25 fractions) and a megavoltage cone-beam computed tomography (MV-CBCT), taken for weekly image verification, and assess the risk of radiation-induced contralateral breast cancer.
Methods And Materials:
Organ doses were measured with thermoluminescent dosimeters placed strategically within a female anthropomorphic phantom. The risk of radiation-induced secondary cancer of the contralateral breast was estimated from these values using excess absolute risk and excess relative risk models.
Results:
The effective dose from a MV-CBCT (8-monitor units) was 35.9 ± 0.2 mSv. Weekly MV-CBCT imaging verification contributes 0.5% and 17% to the total ipsilateral and contralateral breast dose, respectively. For a woman irradiated at age 50 years, the 10-year postirradiation excess relative risk was estimated to be 0.8 and 0.9 for treatment alone and treatment plus weekly MV-CBCT imaging, respectively. The 10-year postirradiation excess absolute risk was estimated to be 4.7 and 5.6 per 10,000 women-years.
Conclusions:
The increased dose and consequent radiation-induced second cancer risk as calculated by this study introduced by the imaging verification protocols utilizing MV-CBCT in breast radiation therapy must be weighed against the benefits of more accurate treatment. As additional image verification becomes more common, it is important that data be collected in regard to long-term malignancy risk.