The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Comparison between bougies and stylets for simulated tracheal intubation with the C-MAC D-blade videolaryngoscope
Bisanth Batuwitage1, Andrew McDonald, Koji Nishikawa
1From the Aintree University Hospital NHS Foundation Trust (BB, AM, KN, SM, PC), CRUK Liverpool Cancer Trials Unit, University of Liverpool (DL) and Centre for Simulation and Patient Safety NHS North West (SM), Liverpool, UK.
Background:
The C-MAC D-blade is a new, highly angulated, videolaryngoscope blade designed for use in patients with difficult airways. Directing a tracheal tube into the trachea with any indirect viewing laryngoscope can be difficult, even with a good view of the laryngeal inlet.
Objective:
To determine which introducing strategy is most suitable for use with the C-MAC D-blade videolaryngoscope.
Design:
Observational manikin study.
Setting:
Centre for Simulation and Patient Safety.
Participants:
Twenty-four anaesthetists of at least 12 months' of experience.
Interventions:
Six tracheal tube introducer strategies (no tracheal tube introducer (TX); hockey stick stylet (SH); Gliderite stylet (SG); bougie with tube loaded distally, near its curved tip (BD); bougie with tube loaded proximally (BP); bougie unloaded until tracheal placement (BU)) for each of two laryngoscopy settings (easy and difficult) in a SimMan 3G manikin. Two intubation attempts, with a maximum time allowance of 60 s each, were allowed for all laryngoscopy setting/introducer combinations.
Main Outcome Measures:
Time to intubate in seconds. Secondary outcome was overall subjective difficulty using a visual analogue scale.
Results:
In the easy laryngoscopy setting, time to intubation was [median (interquartile range): SH 8.5 sec (7 to 11); SG 10 (8 to 11.5); BD 11 (10 to 12.5); TX 11 (7 to 31.5); BP 12 (11 to 13.5); BU 13 (11 to 14.5). There was no evidence of an overall difference in introducer strategies for time to intubate (P < .025) with SG and SH found to be favourable when compared with BU, and SH was also favourable when compared with BP. In the difficult laryngoscopy setting, time to intubation was: SG 11.5 (10 to 17.5); SH 14 (12 to 22); BD 15.5 (12 to 23.5); BU 16.5 (14 to 21); BP 16.5 (15.5 to 20.5); TX 60 (26.5 to 60). There was evidence of an overall difference in introducer strategies for time to intubate (P < 0.025) with all introducers found to be favourable compared with TX. SG was found to be favourable when compared with BU and BP. In groups TX and SH, anaesthetists failed to intubate in 13 of 24 cases and 1 of 24 cases, respectively. The visual analogue scores tended to reflect intubation times, more so in the difficult setting.
Conclusion:
The differences in time to intubate in both the easy and difficult laryngoscopy settings were minor for the different introducer strategies studied. The stylet introducers tended to take less time, but this did not result in important statistical differences. When using the C-MAC D-blade videolaryngoscope for simulated tracheal intubation, stylets and bougies performed in a similar fashion.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...