Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Biometry and intraocular lens power calculation results with a new optical biometry device: comparison with the gold
Godefroy Kaswin1, Antoine Rousseau1, Mohamed Mgarrech1
1From the Service d'Ophtalmologie (Kaswin, Rousseau, Mgarrech, Barreau, Labetoulle), Hôpital Bicêtre, APHP, Université Paris Sud, Le Kremlin-Bicêtre and Laboratoire de Virologie Moléculaire et Structurale CNRS UPR 3296 (Rousseau, Labetoulle), Gif sur Yvette, France.
Purpose:
To evaluate the agreement in axial length (AL), keratometry (K), anterior chamber depth (ACD) measurements; intraocular lens (IOL) power calculations; and predictability using a new partial coherence interferometry (PCI) optical biometer (AL-Scan) and a reference (gold standard) PCI optical biometer (IOLMaster 500).
Setting:
Service d'Ophtalmologie, Hopital Bicêtre, APHP Université, Paris, France.
Design:
Evaluation of a diagnostic device.
Methods:
One eye of consecutive patients scheduled for cataract surgery was measured. Biometry was performed with the new biometer and the reference biometer. Comparisons were performed for AL, average K at 2.4 mm, ACD, IOL power calculations with the Haigis and SRK/T formulas, and postoperative predictability of the devices. A P value less than 0.05 was statistically significant.
Results:
The study enrolled 50 patients (mean age 72.6 years±4.2 SEM). There was a good correlation between biometers for AL, K, and ACD measurements (r=0.999, r=0.933, and r=0.701, respectively) and between IOL power calculation with the Haigis formula (r=0.972) and the SRK/T formula (r=0.981). The mean absolute error (MAE) in IOL power prediction was 0.42±0.08 diopter (D) with the new biometer and 0.44±0.08 D with the reference biometer. The MAE was 0.20 D with the Haigis formula and 0.19 with the SRK/T formula (P=.36).
Conclusion:
The new PCI biometer provided valid measurements compared with the current gold standard, indicating that the new device can be used for IOL power calculations for routine cataract surgery.
Financial Disclosure:
No author has a financial or proprietary interest in any material or method mentioned.