Left and right ventricular lead positions are imprecisely determined by fluoroscopy in cardiac resynchronization
Anders Sommer1, Mads Brix Kronborg2, Bjarne Linde Nørgaard2
1Department of Cardiology, Aarhus University Hospital, Skejby, Brendstrupgaardsvej 100, Aarhus N DK-8200, Denmark a.sommer@dadlnet.dk.
Aims:
Fluoroscopy is the routine method for localizing left ventricular (LV) and right ventricular (RV) lead positions in cardiac resynchronization therapy (CRT). However, the ability of fluoroscopy to determine lead positions in a standard ventricular segmentation is unknown. We aimed to evaluate the accuracy and reproducibility of fluoroscopy to determine LV and RV lead positions in CRT when compared with cardiac computed tomography (CT).
Methods And Results:
Fifty-nine patients undergoing CRT were included. Bi-plane fluoroscopy and cardiac CT were evaluated in all patients. Pacing lead positions were assessed in a standard LV 16-segment model and in a simplistic RV 8-segment model. Four patients with LV lead displacement were excluded from the agreement analysis of LV lead position. Agreement of LV lead position between fluoroscopy and cardiac CT was observed in 19 (35%) patients with fluoroscopy demonstrating a 1-segment and ≥2-segment error in 30 (55%) and 6 (11%) patients, respectively. Agreement of RV lead position was found in 13 (22%) patients with fluoroscopy showing a 1-segment and ≥ 2-segment error in 28 (47%) and 18 (31%) patients, respectively. The interobserver agreement on LV and RV lead positions was poor for fluoroscopy (kappa 0.20 and 0.23, respectively) and excellent for cardiac CT (kappa 0.87 and 0.85, respectively).
Conclusion:
Fluoroscopy is inaccurate and modestly reproducible when assessing LV and RV lead positions in a standard ventricular segmentation when compared with cardiac CT. Cardiac CT should be applied to determine the exact pacing site in future research evaluating the optimal pacing lead position in CRT.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies for Cardiovascular System V: CT
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...