Related Concept Videos

Instrumentation Amplifier01:25

Instrumentation Amplifier

1.3K
An electrocardiography (ECG) machine is an essential piece of medical equipment used to monitor the electrical activity of the heart. It operates by detecting small electrical changes on the skin that result from the depolarization of the heart muscle during each heartbeat. However, these signals are in the microvolt range and can be easily overwhelmed by noise or interference.
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
1.3K
High-Performance Liquid Chromatography: Types of Detectors01:15

High-Performance Liquid Chromatography: Types of Detectors

2.3K
The role of the detectors in High-Performance Liquid Chromatography (HPLC) is to analyze the solutes as they exit from the chromatographic column. The detector recognizes the solute's property and generates corresponding electrical signals, which are converted into a readable graph of the detector's response versus elution time called a chromatogram at the computer. There are several types of HPLC detectors, each with its own advantages and limitations, depending on the analyte...
2.3K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Predicting survival in oral squamous cell carcinoma via integrated analysis of tumor budding and tertiary lymphoid structures.

Frontiers in oncology·2026
Same author

Imaging of two-dimensional ion-beam profiles using a scanning Faraday cup array combined with a 128-channel picoammeter system.

The Review of scientific instruments·2026
Same author

Acoustic measurement methods and spatiotemporal distribution patterns of microbubble spectra in water under artificial aeration conditions.

Ultrasonics·2026
Same author

Treatment options for long head of biceps tendon tenodesis.

Frontiers in surgery·2026
Same author

A deep learning-driven automated treatment planning framework for cervical cancer patients treated with volumetric modulated arc therapy.

Radiation oncology (London, England)·2026
Same author

Guiding Fast Ion Beam by Suppressing Secondary Ions.

Physical review letters·2026

Related Experiment Video

Updated: May 1, 2026

Probing Nicotinic Acetylcholine Receptor Function in Mouse Brain Slices via Laser Flash Photolysis of Photoactivatable Nicotine
10:48

Probing Nicotinic Acetylcholine Receptor Function in Mouse Brain Slices via Laser Flash Photolysis of Photoactivatable Nicotine

Published on: January 25, 2019

10.6K

Note: A timing micro-channel plate detector with backside fast preamplifier.

Wei Wang1, Deyang Yu1, Rongchun Lu1

  • 1Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China.

The Review of Scientific Instruments
|April 3, 2014
PubMed
Summary
This summary is machine-generated.

A new micro-channel plate detector with a backside preamplifier was developed to prevent signal distortion. This timing detector achieved a fast ~2 ns rise time with low background noise.

More Related Videos

A Wireless, Bidirectional Interface for In Vivo Recording and Stimulation of Neural Activity in Freely Behaving Rats
10:41

A Wireless, Bidirectional Interface for In Vivo Recording and Stimulation of Neural Activity in Freely Behaving Rats

Published on: November 7, 2017

14.8K
Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples
07:01

Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples

Published on: June 9, 2016

9.2K

Related Experiment Videos

Last Updated: May 1, 2026

Probing Nicotinic Acetylcholine Receptor Function in Mouse Brain Slices via Laser Flash Photolysis of Photoactivatable Nicotine
10:48

Probing Nicotinic Acetylcholine Receptor Function in Mouse Brain Slices via Laser Flash Photolysis of Photoactivatable Nicotine

Published on: January 25, 2019

10.6K
A Wireless, Bidirectional Interface for In Vivo Recording and Stimulation of Neural Activity in Freely Behaving Rats
10:41

A Wireless, Bidirectional Interface for In Vivo Recording and Stimulation of Neural Activity in Freely Behaving Rats

Published on: November 7, 2017

14.8K
Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples
07:01

Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples

Published on: June 9, 2016

9.2K

Area of Science:

  • Nuclear instrumentation
  • Particle detection technologies

Background:

  • Signal distortion in micro-channel plate detectors can limit timing precision.
  • Traditional preamplifier configurations may introduce signal degradation.

Purpose of the Study:

  • To develop a timing micro-channel plate detector with a backside preamplifier to minimize signal distortion.
  • To characterize the performance of the novel detector system.

Main Methods:

  • Designed and constructed a micro-channel plate detector with an integrated backside double-channel fast preamplifier.
  • Described the mechanical and electronic architecture of the detector system.
  • Tested the detector's performance using an Americium-241 (241Am) alpha-source.

Main Results:

  • Achieved a signal rise time of approximately 2 nanoseconds (~2 ns).
  • Measured an output background noise level of 4 millivolts root-mean-square (4 mV(rms)).
  • Successfully mitigated signal distortion during propagation from the anode to the preamplifier.

Conclusions:

  • The developed backside preamplifier design effectively preserves signal integrity in micro-channel plate detectors.
  • The detector system demonstrates excellent timing resolution suitable for high-speed applications.
  • This advancement offers improved performance for particle detection and timing measurements.