A new calibration concept for a reproducible quantitative detection based on SERS measurements in a microfluidic
Evelyn Kämmer1, Konstanze Olschewski, Thomas Bocklitz
1Institute of Physical Chemistry and Abbe Center of Photonics, Friedrich Schiller University, Helmholtzweg 4, 07743 Jena, Germany. juergen.popp@uni-jena.de.
Abstract:
This study demonstrates a new concept of calibrating surface enhanced Raman scattering (SERS) intensities without using additional substances as an internal standard and explores factors such as laser fluctuation and different Ag substrates, which affect the results of quantitative analyses based on SERS. To demonstrate the capabilities of the concept, the model analyte adenine has been chosen. A lab-on-a-chip device is applied for the measurements to guarantee consistent data recording. In order to simulate varied measuring conditions, two varying silver colloids (batch 1 and 2) are utilized as a SERS substrate and two different laser power levels (25 or 55 mW) are applied on the sample. A concentration gradient was generated which allows the use of the analyte itself for the correction of the resulting SERS spectra regarding intensity deviations caused by different ambient conditions. In doing so, a vast improvement in the quantification using SERS, especially in view of the comparability, reproducibility and repeatability, is demonstrated.
![Quantitative SERS Detection of Uric Acid via Formation of Precise Plasmonic Nanojunctions within Aggregates of Gold Nanoparticles and Cucurbit[n]uril](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F61682.jpg&w=3840&q=50)