First experience of intraoperative echocardiography-guided optimization of cardiac resynchronization therapy delivery
Ghassan Moubarak1, Philippe Ritter2, Jean-Claude Daubert3
1Department of Pacing and Interventional Electrophysiology, Groupe Hospitalier Paris Saint-Joseph, 185, rue Raymond-Losserand, 75014 Paris, France.
Background:
Insufficient correction of mechanical dyssynchrony is a cause of non-response to cardiac resynchronization therapy (CRT).
Aims:
To determine if CRT delivery could be optimized during the implantation procedure by choosing the number and location of pacing sites using echocardiography guidance.
Methods:
In patients with a QRS ≥ 150 ms or a QRS < 150 ms and criteria for mechanical dyssynchrony, the objective of the implantation procedure was to shorten the left pre-ejection interval (LPEI), measured online, by at least 10 ms compared with standard biventricular configuration, by moving the right ventricular (RV) lead at different locations and, if necessary, by adding a second RV lead.
Results:
Ninety-one patients (70 men; mean age 73 ± 10 years; left ventricular [LV] ejection fraction 29 ± 10%) were included. The final pacing configuration was standard biventricular in 15 (17%) patients, optimized biventricular in 22 (24%) and triple-site ventricular in 54 (59%). LPEI was shortened by ≥ 10 ms compared with standard biventricular stimulation in 73 (80%) patients. Compared with standard biventricular pacing, the final optimized pacing configuration improved global intraventricular synchrony (decreasing LPEI from 158 ± 36 ms to 134 ± 29 ms; P<0.001), LV systolic efficiency (decreasing LPEI/LV ejection time from 0.58 ± 0.18 to 0.46 ± 0.13; P<0.001) and LV filling (increasing LV filling time/RR from 44 ± 8% to 47 ± 7%; P<0.001) and decreased mitral valve regurgitation.
Conclusion:
Intraoperative echocardiography-guided placement of RV lead(s) during CRT implantation is feasible and acutely improves LV synchrony compared with standard biventricular stimulation.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
08:42Cox-Maze IV Procedure Concomitant with Valvular Surgery In Situs Inversus Dextrocardia: A Single-Center Experience in China
Published on: February 11, 2022
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Cardiopulmonary Resuscitation III: AED Use
Cardiac Catheterization I: Pre-Procedure Overview