Small-diameter titanium Grade IV and titanium-zirconium implants in edentulous mandibles: three-year results from a
Marc Quirynen1, Bilal Al-Nawas2, Henny J A Meijer3
1Catholic University Leuven, Leuven, Belgium.
Objective:
The aim of this study was to compare crestal bone-level changes, soft tissue parameters and implant success and survival between small-diameter implants made of titanium/zirconium (TiZr) alloy or of Grade IV titanium (Ti) in edentulous mandibles restored with removable overdentures.
Materials And Methods:
This was a randomized, controlled, double-blind, split-mouth multicenter clinical trial. Patients with edentulous mandibles received two Straumann bone-level implants (diameter 3.3 mm), one of Ti Grade IV (control) and one of TiZr (test), in the interforaminal region. Implants were loaded after 6-8 weeks and removable Locator-retained overdentures were placed within 2 weeks of loading. Modified plaque and sulcus bleeding indices, radiographic bone level, and implant survival and success were evaluated up to 36 months.
Results:
Of 91 treated patients, 75 completed the three-year follow-up. Three implants were lost (two control and one test implant). The survival rates were 98.7% and 97.3%, and the mean marginal bone level change was -0.78 ± 0.75 and -0.60 ± 0.71 mm for TiZr and Ti Grade IV implants. Most patients had a plaque score of 0 or 1 (54% for test and 51.7% for control), and a sulcus bleeding score of 0 (46.1% for test and 44.9% for control). No significant differences were found between the two implant types for bone-level change, soft tissue parameters, survival and success.
Conclusions:
After 36 months, similar outcomes were found between Ti Grade IV and TiZr implants. The results confirm that the results seen at 12 months continue over time.