Neointimal hyperplasia after stent placement across size-discrepant vessels in an animal study
Hisayuki Cho1, Mineyoshi Nango, Yukimasa Sakai
1Department of Radiology, Japan Community Health care Organization Osaka Hospital, 4-2-78 Fukushima, Fukushima-ku, Osaka, 553-0003, Japan.
Purpose:
To examine differences in neointimal hyperplasia (NIH) after placing a self-expanding (SE) stent across size-discrepant vessels.
Materials And Methods:
The subjects were 6 beagles, and the target vessels were the abdominal aorta and the external iliac artery (EIA). A nitinol SE stent was placed which was normal-sized in the aorta and oversized in the EIA. Angiography and intravascular ultrasound (IVUS) were performed immediately and after 1 and 3 months; histopathologic examinations were then performed. Furthermore, the chronic outward force (COF) on the same type of stent was investigated in vitro.
Results:
On IVUS, thickened intima was seen on the EIA at 1 month (5.1 ± 4.2 mm(2)) and at 3 months (7.8 ± 2.5 mm(2)). For the aorta, thickening of the intima was negligible at any time. Histopathologically, the percentage of the vessel obliterated by NIH was significantly greater on the iliac side than on the aortic side (33.2 ± 10.4 vs. 13.4 ± 4.4 %). The COF exerted when stent diameter reached that of the EIA and the aorta was 0.73 and 0.17 N/mm(2), respectively.
Conclusions:
When a non-tapered stent is placed in vessels with a large discrepancy in diameter, attention must be paid to increased NIH in the oversized side.