Titanium alloy stem as a cause for adverse reaction to metal debris after bipolar hemiarthroplasty
Related Experiment Video
Updated: May 1, 2026

Potentiodynamic Corrosion Testing
Published on: September 4, 2016
Masaaki Sakamoto1, Hitoshi Watanabe1, Hidetaka Higashi2
1Department of Orthopaedic Surgery, Chiba Aoba Municipal Hospital, 1273-2 Aoba-cho, Chuo-ku, Chiba 260-0852, Japan.
Abstract:
A 68-year-old male with failure of bipolar hemiarthroplasty consistent with adverse reaction to metal debris (ARMD) who presented with a painful cystic lesion and lower extremity swelling was encountered. However, revision surgical findings showed no apparent cause of ARMD previously described in the literature, such as corrosion at the head-neck junction and articular abrasion. Therefore, it was difficult to make a definite diagnosis of failure secondary to ARMD, which consequently led to the decision to perform two-stage revision procedure, though the stem was firmly fixed. Postoperative analysis in the retrieval tissues showed that the metal debris mainly originated from the titanium alloy stem itself. Although this is a very rare case, one should be aware that even the well-fixed femoral components themselves have the potential to be the cause of ARMD.
More Related Videos
09:08Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018