Pressure and flow waveform characteristics of eight high-frequency oscillators
Edward R Harcourt1, Jubal John, Peter A Dargaville
11Neonatology, Royal Children's Hospital, Parkville, VIC, Australia. 2Clinical Technology Service, Royal Children's Hospital, Parkville, VIC, Australia. 3Neonatal Research, Royal Women's Hospital, Parkville, VIC, Australia. 4Neonatal Research, Murdoch Children's Research Institute, Parkville, VIC, Australia. 5Department of Paediatrics, Royal Hobart Hospital, Hobart, TAS, Australia. 6Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano University, Milano, Italy. 7Departments Paediatrics and Obstetrics and Gynaecology, University of Melbourne, Melbourne, VIC, Australia.
Objectives:
The differences in performance of early generation high-frequency oscillators have been attributed to their distinct pressure and flow waveforms. Recently, five new oscillators have been commercially released. The objective of this study was to characterize the pressure and flow waveforms of eight commercially available oscillators.
Design:
In vitro benchtop study.
Setting:
Tertiary pediatric teaching hospital.
Interventions:
Eight oscillators were evaluated using a test lung; mean airway pressure 10 and 20 cm H2O; frequencies 5, 10, and 15 Hz; pressure amplitude 30 cm H2O (or equivalent); compliance 1.0 mL/cm H2O; and endotracheal tube 3.5 mm. Ventilators tested were Sensormedics 3100A and B (Carefusion), SLE5000 (SLE), Fabian (Acutronic), Leonie+ (Heinen+Löwenstein), Sophie (Stephan), and VN500 and Babylog 8000 (Dräger).
Measurements And Main Results:
Pressure (airway opening, at oscillator and within the test lung) and airway opening flow waveforms were recorded. Airway opening waveforms were characterized by type (square or sine) and by determining power spectral density analysis. The Sensormedics A and B and the SLE5000 delivered square waves; all other oscillators generated sine waves. Sensormedics, the SLE5000, and the Sophie had a characteristic inspiratory slope (incisura). The pressure waveform within the test lung was a sine wave for all oscillators. Oscillators with square waves or an inspiratory incisura exhibited the highest number of nonfundamental frequency components on power spectral density analysis, suggesting more complex harmonic waveforms with potentially greater transmissive power to the lungs. At frequencies of 5 and 10 Hz, all ventilators, except Babylog 8000, generated airway pressure amplitudes greater than 28.6 cm H2O and tidal volumes greater than 6 mL at the airway opening.
Conclusions:
Current high-frequency oscillators deliver different waveforms. As these may result in variable clinical performance, operators should be aware that these differences exist.
Related Concept Videos
Forced Oscillations
Effective Value of a Periodic Waveform
The effective value of a periodic current represents the direct current (DC) that conveys the same average power to a resistor as the periodic current itself. This concept is crucial when assessing AC circuits. To determine the...
Sound as Pressure Waves
The pressure fluctuation depends on the difference in displacements between the successive points in the...
Oscillations In An LC Circuit
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Characteristics of OpAmp