An in vitro evaluation of the impact of eccentric deployment on transcatheter aortic valve hemodynamics

Paul S Gunning1, Neelakantan Saikrishnan, Laoise M McNamara

  • 1Biomechanics Research Centre, Department of Biomedical Engineering, National University of Ireland Galway, Galway, Ireland.

Insights

Eccentric deployment of Transcatheter Aortic Valve Replacement (TAVR) stents causes turbulent blood flow and may increase blood cell damage. This occurs due to altered flow characteristics not detected by standard hemodynamic evaluation alone.

Area of Science:

  • Cardiovascular Research
  • Biomedical Engineering
  • Fluid Dynamics

Background:

  • Aortic stenosis involves calcium deposits on native valves, potentially causing non-concentric Transcatheter Aortic Valve Replacement (TAVR) stent expansion.
  • Eccentric TAVR deployment may impact blood flow dynamics and cellular integrity.

Purpose of the Study:

  • To investigate if eccentric TAVR deployment leads to turbulent blood flow and blood cell damage.
  • To compare hemodynamic parameters and downstream flow characteristics between circular and eccentric TAVR deployments.

Main Methods:

  • Utilized Particle Image Velocimetry (PIV) to analyze fluid velocity fields.
  • Quantified shear stress and turbulent kinetic energy downstream of TAVRs in circular and eccentric configurations.
  • Simulated TAVR deployment in representative circular and eccentric orifices.

Main Results:

  • Eccentric TAVR deployment resulted in asymmetric systolic jet formation.
  • Increased shear stresses (119 N/m² vs. 97 N/m²) and turbulence intensity (230 N/m² vs. 180 N/m²) were observed downstream of eccentrically deployed TAVRs.
  • Effective orifice area and mean transvalvular pressure gradient showed minimal differences between circular and eccentric deployments.

Conclusions:

  • Eccentric TAVR deployment alters blood flow dynamics, creating turbulence and elevated shear stress.
  • These altered flow characteristics may increase the hemolytic potential of TAVR valves.
  • Hemodynamic evaluation alone may not fully capture the risks associated with eccentric TAVR deployment.