Updated: May 1, 2026

Treatment of Ankle Osteoarthritis with Total Ankle Replacement Through a Lateral Transfibular Approach
09:01

Treatment of Ankle Osteoarthritis with Total Ankle Replacement Through a Lateral Transfibular Approach

Published on: January 24, 2018

12.0K

Total Ankle Arthroplasty Accuracy and Reproducibility Using Preoperative CT Scan-Derived, Patient-Specific Guides.

Gregory C Berlet1, Murray J Penner2, Sarah Lancianese3

  • 1Orthopedic Foot & Ankle Center, Westerville, OH, USA gberlet@gmail.com.

Foot & Ankle International
|April 11, 2014
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Long-term risk of subtalar joint arthritis after stemmed total ankle replacement.

The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons·2026
Same author

Short-term risk of subtalar joint arthritis after stemmed total ankle replacement.

The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons·2025
Same author

Anterior Approach Total Ankle Arthroplasty with Patient-Specific Cut Guides.

JBJS essential surgical techniques·2025
Same author

Tranexamic acid and its outcomes in foot and ankle surgical procedures: A meta-analysis.

The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons·2025
Same author

Corrigendum to "A Study of Tibial Cyst Formation in Modular Stemmed Total Ankle Arthroplasty: Exploring a Possible Relationship to Smooth and Porous Coating on the Stem Segments" [Journal of Foot and Ankle Surgery 62 (2023) 756-763].

The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons·2025
Same author

A Study of Tibial Cyst Formation in Modular Stemmed Total Ankle Arthroplasty: Exploring a Possible Relationship to Smooth and Porous Coating on the Stem Segments.

The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons·2023

This study shows that patient-specific guides for total ankle arthroplasty (TAA) offer reliable implant placement. The technology achieved less than 2 degrees of deviation, improving accuracy for TAA procedures.

Area of Science:

  • Orthopaedic Surgery
  • Biomedical Engineering
  • Medical Imaging

Background:

  • Preoperative navigation offers benefits like patient-specific alignment in total knee arthroplasty.
  • This technology was newly applied to total ankle arthroplasty (TAA).
  • The study assessed the repeatability of patient-specific guide placement for tibia and talus components in TAA.

Purpose of the Study:

  • To evaluate the repeatability of patient-specific guide placement in TAA.
  • To determine the deviation between the preoperative plan and actual implant placement.
  • To assess the accuracy of custom guides for achieving patient-specific ankle alignment.

Main Methods:

  • 15 cadaveric lower extremity limbs underwent CT scans and 3D model creation.
  • Virtual TAA was performed, and custom guides were engineered based on patient-specific anatomy.
Keywords:
arthritispreoperative navigation systemtotal ankle arthroplasty

More Related Videos

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
07:45

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report

Published on: August 4, 2022

2.9K

Related Experiment Videos

Last Updated: May 1, 2026

Treatment of Ankle Osteoarthritis with Total Ankle Replacement Through a Lateral Transfibular Approach
09:01

Treatment of Ankle Osteoarthritis with Total Ankle Replacement Through a Lateral Transfibular Approach

Published on: January 24, 2018

12.0K
The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
07:45

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report

Published on: August 4, 2022

2.9K
  • Orthopaedic surgeons placed the guides multiple times to assess variability, with final implant positions verified by CT scans and compared to the preoperative plan.
  • Main Results:

    • Intraobserver variation for tibia and talus guide placement was minimal across all rotational degrees of freedom (e.g., <1.2 degrees for internal/external rotation).
    • Average deviation between planned and actual implant placement was less than 2 degrees and 1.4 mm in all tested specimens.
    • Reproducibility of tibial component placement was within 2 degrees of the intended target.

    Conclusions:

    • Preliminary data indicate that preoperative navigation with custom guides ensures reliable and reproducible TAA implant placement.
    • The achieved accuracy (less than 2 degrees deviation) surpasses that of traditional instrumentation in similar studies.
    • This technology shows promise for becoming a benchmark for future total ankle arthroplasty procedures.