Rapid identification of falsely elevated serum cardiac troponin I values in a stat laboratory
Objectives:
To establish a rapid method for the stat laboratory to identify falsely elevated serum cardiac troponin I (cTnl) values caused by endogenous interferences.
Methods:
Serum specimens with elevated cTnI were recentrifuged and cTnI was measured again. If the difference between first and second cTnI value was less than +/- 30%, the specimen was diluted 1:2, 1:5, and 1:10, and the cTnI was measured in the diluted specimens. If the cTnI concentrations did not change linearly with dilution, the specimen was referred to an outside lab for re-analysis of cTnI by a different immunochemical method.
Results:
Troponin I was elevated in 132 serum specimens. In 18 specimens, we attributed the elevation to the presence of fibrin or microparticles in the serum. Serial dilutions of 3 of the 114 remaining specimens produced nonlinear results. Overall, falsely elevated cTnI results were identified in 1.01% of all measurements.
Conclusions:
Our method, which uses combined measurement of cTnI, creatinine kinase MB isoenzyme (CK-MB), and myoglobin, together with incubation, re-centrifugation, and serial dilution, is a practical method to identify falsely elevated serum cTnI values.
More Related Videos
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome III: Diagnostic Studies
Acute Coronary Syndrome I: Introduction
Myocarditis II: Clinical Features and Diagnostic Tests
Errors occurring during blood pressure monitoring
Several factors...