Platelet function during cardiopulmonary bypass using multiple electrode aggregometry: comparison of centrifugal and

Hiromu Kehara1, Tamaki Takano, Noburo Ohashi

  • 1Department of Cardiovascular Surgery, Shinshu University School of Medicine, Matsumoto, Japan.

Artificial Organs
|April 12, 2014
PubMed

Insights

Centrifugal pumps (CPs) and roller pumps (RPs) showed no significant difference in platelet function during cardiopulmonary bypass (CPB). Initial contact with foreign materials, not pump type, appears to cause platelet dysfunction.

Area of Science:

  • Cardiovascular Surgery
  • Biomedical Engineering
  • Hematology

Background:

  • Cardiopulmonary bypass (CPB) utilizes pumps to maintain circulation during surgery.
  • Centrifugal pumps (CPs) are theorized to cause less blood trauma than roller pumps (RPs) due to non-occlusive flow.
  • However, comparative effects on platelet function remain incompletely understood.

Purpose of the Study:

  • To compare the impact of CPs versus RPs on platelet function during CPB.
  • To investigate whether pump type influences platelet activation and aggregation.
  • To determine the primary drivers of platelet dysfunction during CPB.

Main Methods:

  • Ten swine underwent 3-hour CPB, with five using CPs and five using RPs.
  • Platelet function was assessed using multiple electrode aggregometry (MEA) at various time points.
  • Platelet activation was stimulated by adenosine diphosphate (ADP), arachidonic acid (AA), and thrombin receptor-activating protein 6 (TRAP).

Main Results:

  • MEA showed no significant differences in platelet function between CP and RP groups at any time point.
  • Platelet aggregation decreased significantly in both groups with ADP and AA stimulation, recovering post-CPB.
  • Fibrinogen, platelet factor 4 (PF4), and β-thromboglobin (β-TG) levels were similar between groups post-CPB.

Conclusions:

  • Platelet function assessed by MEA does not significantly differ between CPs and RPs during CPB.
  • Platelet dysfunction appears primarily related to initial contact with foreign materials, not the specific pump type.
  • Further research may elucidate specific mechanisms of pump-induced blood trauma on platelets.