Related Experiment Video
Updated: May 1, 2026

Ultrasensitive Detection of Biomarkers by Using a Molecular Imprinting Based Capacitive Biosensor
Published on: February 16, 2018
Liquid crystal and gold nanoparticles applied to electrochemical immunosensor for cardiac biomarker
Eduardo Zapp1, Eduard Westphal2, Hugo Gallardo2
1Laboratory of Biosensors, Department of Chemistry, Federal University of Santa Catarina, 88040-900 Florianópolis, Santa Catarina, Brazil.
Abstract:
A label-free electrochemical immunosensor based on the ionic liquid crystal (E)-1-decyl-4-[(4-decyloxyphenyl)diazenyl]pyridinium bromide (Br-Py) coated on a glassy carbon electrode (GCE) for the quantitative detection of myoglobin (Mb), a cardiac marker for acute myocardial infarction, is reported herein for the first time. The monoclonal anti-myoglobin antibody (ab-Mb) was covalently immobilized using glyoxal on a film of polyethyleneimine-coated gold nanoparticles (AuNP-PEI). The proposed method for Mb detection is based on voltammetric suppression of the Br-Py signal when the immunosensor was incubated with Mb antigen. The electrochemical performance of the Mb immunosensor was studied by electrochemical impedance spectroscopy, and cyclic and square-wave voltammetry. Under the optimal conditions, the proposed immunosensor shows a good linear relationship between the electrochemical inhibition response and the concentration of Mb over the range of 9.96-72.8 ng mL(-1) with a detection limit of 6.29 ng mL(-1). The results obtained indicate that the proposed immunosensor provides good sensitivity and simple operation for detecting acute myocardial infarction with Mb as a biomarker.