DLS 5.0--the biomechanical effects of dynamic locking screws
Stefan Döbele1, Michael Gardner2, Steffen Schröter1
1BG Trauma Center, Eberhard Karls Universitaet Tuebingen, Tübingen, Germany.
Introduction:
Indirect reduction of dia-/metaphyseal fractures with minimally invasive implant application bridges the fracture zone in order to protect the soft-tissue and blood supply. The goal of this fixation strategy is to allow stable motion at the fracture site to achieve indirect bone healing with callus formation. However, concerns have arisen that the high axial stiffness and eccentric position of locked plating constructs may suppress interfragmentary motion and callus formation, particularly under the plate. The reason for this is an asymmetric fracture movement. The biological need for sufficient callus formation and secondary bone healing is three-dimensional micro movement in the fracture zone. The DLS was designed to allow for increased fracture site motion. The purpose of the current study was to determine the biomechanical effect of the DLS_5.0.
Methods:
Twelve surrogate bone models were used for analyzing the characteristics of the DLS_5.0. The axial stiffness and the interfragmentary motion of locked plating constructs with DLS were compared to conventional constructs with Locking Head Screws (LS_5.0). A quasi-static axial load of 0 to 2.5 kN was applied. Relative motion was measured.
Results:
The dynamic system showed a biphasic axial stiffness distribution and provided a significant reduction of the initial axial stiffness of 74.4%. Additionally, the interfragmentary motion at the near cortex increased significantly from 0.033 mm to 0.210 mm (at 200N).
Conclusions:
The DLS may ultimately be an improvement over the angular stable plate osteosynthesis. The advantages of the angular stability are not only preserved but even supplemented by a dynamic element which leads to homogenous fracture movement and to a potentially uniform callus distribution.
More Related Videos
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Screw: Problem Solving
To evaluate the...
Deformation in a Circular Shaft
Journal Bearings
To better understand the concept of journal bearings, consider a rope winch with dry or...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...