A novel hydrogen peroxide biosensor based on hemoglobin-collagen-CNTs composite nanofibers
1Key Laboratory of Green Chemical Engineering and Technology of College of Heilongjiang Province, College of Chemical and Environmental Engineering, Harbin University of Science and Technology, Harbin, 150080, China.
Abstract:
In this paper, carbon nanotubes (CNTs) were successfully incorporated in the composite composed of hemoglobin (Hb) and collagen using co-electrospinning technology. The formed Hb-collagen-CNTs composite nanofibers possessed distinct advantage of three-dimensional porous structure, biocompatibility and excellent stability. The Hb immobilized in the electrospun nanofibers retained its natural structure and the heterogeneous electron transfer rate constant (ks) of the direct electron transfer between Hb and electrodes was 5.3s(-1). In addition, the electrospun Hb-collagen-CNTs nanofibers modified electrodes showed good electrocatalytic properties toward H2O2 with a detection limit of 0.91μM (signal-to-noise ratio of 3) and the apparent Michaelis-Menten constant (Km(app)) of 32.6μM.