Tuning a Parallel Segmented Flow Column and Enabling Multiplexed Detection
Published on: December 15, 2015
Band broadening along gradient reversed phase columns: a potential gain in resolution factor
Fabrice Gritti1, Georges Guiochon1
1Department of Chemistry, University of Tennessee, Knoxville, TN 37996-1600, USA.
Abstract:
The advantages of using columns packed with stationary phases having a composition gradient so that retention factors increase toward the outlet and eluting them with an isocratic mobile phase may provide increased peak resolution in liquid chromatography. This approach is discussed from a theoretical viewpoint. The retention factor is assumed to increase linearly along the column. The peak width can be predicted under three different conditions: (1) the possible band compression is neglected (the Giddings model); (2) band compression is taken into account (the Poppe model); and (3) both band compression and extra-column effects are considered in the calculations. The impact of a stationary phase gradient on the resolution of a pair of compound difficult to separate (constant selectivity α=1.05) is illustrated for a 3mm×100mm silica monolithic column of the second generation (porosity ϵt=0.85, plate height H=6μm) operated with a new generation of liquid chromatograph (pre- and post-column volume variances of 10 and 5μL(2), respectively). The results show that the resolution factor RS becomes optimum for a specific positive value of the retention gradient along the column. This optimum depends on the retention factor at the column inlet.
Related Concept Videos
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Principles Of Column Chromatography