Non-invasive multiparametric qBOLD approach for robust mapping of the oxygen extraction fraction
Sebastian Domsch1, Moritz B Mie1, Frederik Wenz2
1Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Germany.
Introduction:
The quantitative blood oxygenation level-dependent (qBOLD) method has not become clinically established yet because long acquisition times are necessary to achieve an acceptable certainty of the parameter estimates. In this work, a non-invasive multiparametric (nimp) qBOLD approach based on a simple analytical model is proposed to facilitate robust oxygen extraction fraction (OEF) mapping within clinically acceptable acquisition times by using separate measurements.
Methods:
The protocol consisted of a gradient-echo sampled spin-echo sequence (GESSE), a T2-weighted Carr-Purcell-Meiboom-Gill (CPMG) sequence, and a T2(*)-weighted multi-slice multi-echo gradient echo (MMGE) sequence. The GESSE acquisition time was less than 5 minutes and the extra measurement time for CPMG/MMGE was below 2 minutes each. The proposed nimp-qBOLD approach was validated in healthy subjects (N = 5) and one patient.
Results:
The proposed nimp-qBOLD approach facilitated more robust OEF mapping with significantly reduced inter- and intra-subject variability compared to the standard qBOLD method. Thereby, an average OEF in all subjects of 27±2% in white matter (WM) and 29±2% in gray matter (GM) using the nimp-qBOLD method was more stable compared to 41±10% (WM) and 46±10% (GM) with standard qBOLD. Moreover, the spatial variance in the image slice (i.e. standard deviation divided by mean) was on average reduced from 35% to 25%. In addition, the preliminary results of the patient are encouraging.
Conclusion:
The proposed nimp-qBOLD technique provides a promising tool for robust OEF mapping within clinically acceptable acquisition times and could therefore provide an important contribution for analyzing tumors or monitoring the success of radio and chemo therapies.
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...