Related Experiment Video
Updated: May 1, 2026

Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
Post-approval study of a highly pulsed, low-shear-rate, continuous-flow, left ventricular assist device, EVAHEART: a
Satoshi Saito1, Kenji Yamazaki1, Tomohiro Nishinaka1
1Department of Cardiovascular Surgery, Tokyo Women's Medical University, Tokyo.
Background:
The EVAHEART left ventricular assist device was approved in 2010 by the Japanese Pharmaceuticals and Medical Devices Agency (PMDA) for bridge to heart transplantation (BTT). However, its effectiveness has not been evaluated since approval. In this study we evaluated the EVAHEART device in a commercial setting in Japan.
Methods:
Ninety-six consecutive patients enrolled in the Japanese Registry for Mechanically Assisted Circulatory Support (J-MACS), who were listed for transplant or likely to be listed and who received an EVAHEART device, were enrolled from 2011 to 2013 at 14 Japanese centers. Patients' survival rates, adverse events and quality-of-life data were obtained from the J-MACS Registry.
Results:
Patients' median age was 43 years (85% male). The Interagency Registry for Mechanically Assisted Circulatory Support profiles revealed 12 patients in Level 1, 45 in Level 2, 37 in Level 3 and 1 in Level 4. The mean support duration was 384.7 days, with a cumulative duration of 101.2 years. The Kaplan-Meier survival rate during support was 93.4% at 6 months, 87.4% at 1 year and 87.4% at 2 years. Seventy-seven patients (80.2%) currently remain on support, 7 received a transplant and 10 died during support. Major adverse events included drive-line infection (14.6%) and neurologic events such as ischemic stroke (17.7%), hemorrhage (13.5%), transient ischemic attack (3.1%), pump thrombosis (1%) and hemolysis (1%). There was no gastrointestinal (GI) bleeding or right heart failure requiring right ventricular assist device (RVAD). There was no pump exchange due to mechanical failure.
Conclusions:
The EVAHEART device provides safe, reliable and long-term circulatory support with improved survival in commercial settings of BTT in Japan, where the transplant waiting period is much longer. Incidences of GI bleeding, hemolysis, right ventricular failure, device thrombosis and mechanical failure were extremely rare in patients on EVAHEART devices.
More Related Videos
07:39Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
05:18Implantation of Left Ventricular Assist Device (LVAD) in Juvenile Landrace Swine: A LVAD Implantation Model of Pediatric Heart Failure
Published on: January 16, 2026
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Heart Failure II: Pathophysiology