Effectiveness of low rate fluoroscopy at reducing operator and patient radiation dose during transradial coronary
Eltigani Abdelaal1, Guillaume Plourde1, Jimmy MacHaalany1
1Quebec Heart-Lung Institute, Laval University, Quebec City, Quebec, Canada.
Objectives:
This study sought to determine the efficacy of low rate fluoroscopy at 7.5 frames/s (FPS) versus conventional 15 FPS for reduction of operator and patient radiation dose during diagnostic coronary angiography (DCA) and percutaneous coronary intervention (PCI) via the transradial approach (TRA).
Background:
TRA for cardiac catheterization is potentially associated with increased radiation exposure. Low rate fluoroscopy has the potential to reduce radiation exposure.
Methods:
Patients undergoing TRA diagnostic angiography ± ad-hoc PCI were randomized to fluoroscopy at 7.5 FPS versus 15 FPS prior to the procedure. Both 7.5 and 15 FPS fluoroscopy protocols were configured with a fixed dose per pulse of 40 nGy. Primary endpoints were operator radiation dose (measured with dosimeter attached to the left side of the thyroid shield in μSievert [μSv]), patient radiation dose (expressed as dose-area product in Gy·cm(2)), and fluoroscopy time.
Results:
From October 1, 2012 to August 30, 2013, from a total of 363 patients, 184 underwent DCA and 179 underwent PCI. Overall, fluoroscopy at 7.5 FPS compared with 15 FPS was associated with a significant reduction in operator dose (30% relative reduction [RR], p < 0.0001); and in patient's dose-area product (19% RR; p = 0.022). When stratified by procedure type, 7.5 FPS compared with 15 FPS was associated with significant reduction in operator dose during both DCA (40% RR; p < 0.0001) and PCI (28% RR; p = 0.0011). Fluoroscopy at 7.5 FPS, compared with 15 FPS, was also associated with substantial reduction in patients' dose-area product during DCA (26% RR; p = 0.0018) and during PCI (19% RR; p = 0.13). Fluoroscopy time was similar in 7.5 FPS and 15 FPS groups for DCA (3.4 ± 2.0 min vs. 4.0 ± 4.7 min; p = 0.42) and PCI (11.9 ± 8.4 min vs. 13.3 ± 9.7 min; p = 0.57), respectively.
Conclusions:
Fluoroscopy at 7.5 FPS, compared with 15 FPS, is a simple and effective method in reducing operator and patient radiation dose during TRA DCA and PCI.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Cardiac Catheterization III: Left Heart Catheterization