Relationship of beam angulation and radiation exposure in the cardiac catheterization laboratory
Shikhar Agarwal1, Akhil Parashar1, Navkaranbir Singh Bajaj2
1Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio.
Objectives:
The aim of this study was to analyze the relationship between beam angulation and air kerma in a modern cardiac catheterization laboratory.
Background:
Recent reports have identified the merits of reducing radiation scatter, an important determinant of radiation dose in the catheterization laboratory. Radiation scatter is poorly characterized in the context of catheterization laboratories using modern digital equipment. Understanding the principles of dosimetry may reduce the radiation exposure to patients, providers, and medical staff.
Methods:
Prospectively captured radiation data were extracted from a database of 1,975 diagnostic catheterizations (DCs) and 755 percutaneous coronary interventions (PCIs), which included 138,342 fluoroscopic and 35,440 acquisition (cine) sequences. Fluoroscopy and acquisition modes were categorized into tertiles based on the total air kerma measured at a standard reference point. Radiation maps were modeled according to the relative proportion of exposure in each projection.
Results:
Median air kerma during DCs and PCIs was 677 and 2,188 mGy, respectively. Fluoroscopy contributed to 66.3% of total dose during PCIs compared with 39.7% during DCs (p < 0.001). Fluoroscopy was more sensitive to changes in angulation with a rapid increase in total air kerma on small increases in beam angulation. Complex spatial maps were created to study the impact of angulation and other covariates on total air kerma. Besides beam angulation, body surface area was the strongest predictor of the total air kerma.
Conclusions:
This study uniquely describes radiation dosimetry using contemporary equipment in a real-world setting. Extreme angulations were associated with high air kerma values. Fluoroscopy compared with acquisition was more sensitive to changes in angulation, with relatively larger increases in total air kerma with small increases in steepness of the angulation.
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization IV: Nursing Management
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Cardiac Catheterization II: Right Heart Catheterization