Determination of bevantolol in human plasma using liquid chromatography-electrospray ionization tandem mass
Li Ren1, Zheng Wang2, Yiceng Lou2
1College of Pharmacy, Tongji Medical College, Huazhong University of Science and Technology, Wuhan 430030, China.
Abstract:
A liquid chromatography-electrospray ionization tandem mass spectrometry method was established and validated for the determination of bevantolol in human plasma using propranolol as the internal standard. The optimal chromatographic behavior of bevantolol and propranolol was achieved on a Welch Ultimate XB-C18 column (5 μm, 150 mm × 2.1mm, Maryland, USA) with a mobile phase of acetonitrile-water (40:60, v/v) containing 10mM ammonium acetate and 0.1% formic acid. The mass spectrometer was operated in selected reaction monitoring mode using the transition m/z 346.1>165.1 for bevantolol and m/z 260.3>116.1 for propranolol. Sample preparation was carried out through protein precipitation with acetonitrile. The calibration curves were linear over the range of 5.00-1,000 ng/ml. The intra- and inter-day precisions were less than 6.7% and 6.6%, respectively. This method was successfully applied to the bioequivalence study of two kinds of bevantolol hydrochloride tablets in 24 Chinese male volunteers in fasting and postprandial experiment.
More Related Videos
11:25Multi-step Preparation Technique to Recover Multiple Metabolite Compound Classes for In-depth and Informative Metabolomic Analysis
Published on: July 11, 2014
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
Bioequivalence studies: Biowaivers
Equivalence: In Vitro and In Vivo Bioequivalence
Bioequivalence of Drugs: Drugs with Multiple Indications
Measurement of Bioavailability: Pharmacodynamic Methods
Bioequivalence Data: Statistical Interpretation
Measurement of Bioavailability: Pharmacokinetic Methods