Variable dose interplay effects across radiosurgical apparatus in treating multiple brain metastases.
Lijun Ma1, Alan Nichol2, Sabbir Hossain3
1Department of Radiation Oncology, University of California, San Francisco, 505 Parnassus Avenue, Room L-08, San Francisco, CA , 94143, USA. lijunma@radonc.ucsf.edu.
Stereotactic radiosurgery (SRS) platforms show significant differences in normal brain tissue doses. Linear accelerator-based SRS platforms deliver higher doses to normal brain tissue compared to Gamma Knife Perfexion (PFX).
Area of Science:
- Radiosurgery
- Medical Physics
- Radiation Oncology
Background:
- Apparatus dependency significantly impacts normal brain tissue doses in multi-target stereotactic radiosurgery (SRS).
- Investigating inter-target dose interplay across contemporary SRS platforms is crucial for understanding these variations.
Purpose of the Study:
- To determine if inter-target dose interplay effects across different radiosurgical treatment platforms explain observed apparatus-dependent normal brain tissue doses.
- To compare dose delivery and treatment times for various SRS systems.
Main Methods:
- Six institutions planned SRS treatments for subsets of 12 targets using Gamma Knife Perfexion (PFX), CyberKnife, Novalis, and Varian TrueBeam FFF.
- Identical dose-volume constraints were applied across all platforms.
- Treatment plans were developed locally and results analyzed centrally.
Main Results:
- All platforms met dose-volume constraints, with minor differences in structures like the lens.
- Peripheral normal brain tissue doses were lowest with PFX and highest with TrueBeam FFF and CyberKnife.
- Volumes of normal brain receiving 12 Gy were 180-290% higher with TrueBeam FFF, Novalis, and CyberKnife compared to PFX. Treatment times were shortest with TrueBeam FFF.
Conclusions:
- Linear accelerator-based SRS platforms (TrueBeam FFF, Novalis, CyberKnife) deliver higher doses to normal brain tissue than Gamma Knife Perfexion (PFX).
- The volume of normal brain receiving therapeutic doses increases more rapidly per target with linac-based SRS.
- TrueBeam FFF offers the shortest treatment times among the evaluated SRS platforms.