Parameters for quantifying bolus retention with high-resolution impedance manometry
1Department of Medicine, Feinberg School of Medicine, Northwestern University, Chicago, IL, USA.
Background:
This study aimed to develop a methodology for quantifying esophageal bolus retention using a high-resolution esophageal impedance topography (EIT) technique.
Methods:
The ability of impedance to quantify bolus retention was validated by comparison with concurrent fluoroscopic imaging (barium bolus) in 10 healthy subjects. High-resolution impedance manometry (HRIM) studies without fluoroscopy were performed in another 15 healthy subjects to define normal values using saline. HRIM data from each subject were analyzed using a MATLAB program customized for calculating the esophageal impedance integral (EII) prior to the contraction wave front as EII1 and after the contraction as EII2, and presented as a ratio of EII2/EII1, which was compared to the percent of barium areas retained in the esophagus on fluoroscopy determined by a blinded reviewer.
Key Results:
In 93% (37/40) of barium swallows, the results from the EIT method were in agreement with fluoroscopy results with one of three patterns: (1) 25 normal bolus transit, (2) eight bolus stasis, and (3) four retrograde escape or reflux. Three swallows (8%) had slight retention identified by EIT, but no retention detected by fluoroscopy. The correlation between percent of bolus retained in the esophagus detected by fluoroscopy and percent of bolus retention (EII2/EII1) after swallows with EIT method was r = 0.96 (p < 0.001) in supine and r = 0.69 (p < 0.001) in upright position.
Conclusions & Inferences:
The EII ratio (EII2/EII1) is a surrogate for the fraction of retained bolus after a swallow and this metric may be useful in better defining esophageal function.
Related Concept Videos
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Methods for Studying Drug Absorption: In situ
The Doluisio method involves perfusing a prepared segment of a rat's small intestine with a solution of radiolabeled drug and a non-absorbable marker. This helps to differentiate between absorbed and non-absorbed drug concentrations. The intestinal segment is connected at both ends using tubing and syringes,...