Three Different Protocols of Corneal Collagen Crosslinking in Keratoconus: Conventional, Accelerated and Iontophoresis
Published on: November 12, 2015
Spatially heterogeneous corneal mechanical responses before and after riboflavin-ultraviolet-A crosslinking
Joel R Palko1, Junhua Tang1, Benjamin Cruz Perez1
1From the College of Medicine (Palko), the Department of Biomedical Engineering (Tang, Perez, Liu), the Center for Biostatistics (Pan), and the Department of Ophthalmology (Liu), Ohio State University, Columbus, Ohio, USA.
Purpose:
To determine the heterogeneous through-thickness strains in the cornea at physiologic intraocular pressures before and after corneal collagen crosslinking (CXL) using noninvasive ultrasound.
Setting:
Department of Biomedical Engineering, Ohio State University, Columbus, Ohio, USA.
Design:
Experimental study.
Methods:
Sixteen paired canine corneoscleral shells were divided into 2 groups. The CXL group completed a standard CXL protocol using riboflavin-ultraviolet-A (UVA) irradiation. The control group was given an identical treatment except UVA irradiation. Ultrasound scans (at 55 MHz) of the cornea were obtained before and after treatment as the corneoscleral shell was inflated from 5 mm Hg to 45 mm Hg to calculate the distributive through-thickness strains in the cornea. The mean radial and tangential strains of the whole cornea layer, as well as those of the anterior, middle, and posterior thirds of the cornea, were compared before and after treatment in the control group and CXL group using linear mixed models with repeated measures.
Results:
Significant reductions in tangential and radial strains occurred in the CXL group (P=.003 and P=.0025, respectively) but not the control group (P=.08 and P=.63, respectively). The anterior third had the smallest strains in all pretreated corneas (P<.001) and posttreated corneas (CXL group, P=.023; control group, P=.01).
Conclusions:
Ultrasound speckle tracking showed heterogeneous strain distributions through the cornea and confirmed that CXL results in a stiffer corneal response (ie, smaller strains during physiologic loadings). This technique may provide a clinical tool to quantify the biomechanical effects of CXL.
Financial Disclosure:
No author has a financial or proprietary interest in any material or method mentioned.
More Related Videos
12:25Second Harmonic Generation Signals in Rabbit Sclera As a Tool for Evaluation of Therapeutic Tissue Cross-linking TXL for Myopia
Published on: January 6, 2018
05:56Scleral Cross-linking Using Riboflavin and Ultraviolet-A Radiation for Prevention of Axial Myopia in a Rabbit Model
Published on: April 3, 2016