Influence of polyvinyl chloride infusion extension tube on propofol injection pain: a randomised controlled study
1From the Department of Anaesthesiology, Qilu Hospital of Shandong University (QW, NZ, WL), Shandong Quality Inspection Centre for Medical Devices, Jinan, China (YS, YJ).
Background:
Propofol injection pain is a common and unsolved anaesthesia problem.
Objectives:
The present study attempted to confirm that the plasticiser di(2-ethylhexyl) phthalate in polyvinyl chloride (PVC) infusion tubes may increase propofol injection pain by increasing the aqueous propofol concentration.
Design:
A randomised controlled study.
Setting:
University teaching hospital, 1 April to 25 June 2013.
Patients:
One hundred patients scheduled for elective surgery were allocated randomly to the PVC or the control (C) group. The PVC group received a propofol (Diprivan) infusion via a 1-m PVC infusion extension tube, whereas group C received propofol injected directly through the port of the cannula.
Intervention:
After the syringe was loaded with propofol, air was expelled from the tube and the syringe was left standing for 5 min; intravenous propofol 0.5 mg kg was then injected either through the PVC tube or directly into the cannula.
Main Outcome Measure:
A verbal rating scale was used to evaluate the propofol injection pain in both groups. Di(2-ethylhexyl) phthalate and aqueous propofol concentrations were also measured in samples of propofol after simulated injection. To investigate whether the increase in aqueous propofol concentration was caused by leached di(2-ethylhexyl) phthalate, the same amount of di(2-ethylhexyl) phthalate as that measured in the PVC group was added to the samples (group D).
Results:
The incidences of pain in groups PVC and C were 88 and 46%, respectively (P < 0.0001). The di(2-ethylhexyl) phthalate concentration in group PVC (1.01 ± 0.07 μg ml) was greater than that in group C (lower than the detection limit of 0.03 μg ml). No significant difference was found between the aqueous propofol concentrations in groups PVC (25.9 ± 1.8 μg ml) and D (24.4 ± 1.1 μg ml) (P = 0.22), which were significantly higher than that in group C (14.3 ± 1.0 μg ml) (P = 0.079).
Conclusion:
Propofol injection pain is increased by contact with PVC infusion tubing as a result of an increase in aqueous propofol concentration caused by di(2-ethylhexyl) phthalate leaching into the lipid emulsion.
Trial Registration:
chictr.org identifier: ChiCTR-TRC-12003170.
Related Concept Videos
Parenteral Anesthetics: Overview
Local Anesthetics: Clinical Application as Epidural Anesthesia
Since epidural anesthetics can be infused through an epidural catheter, all types of drugs, including short-acting ones, can be administered. Chloroprocaine and lidocaine are examples of short and long-duration anesthetics, respectively. Bupivacaine...
Local Anesthetics: Clinical Application as Intravenous Regional Anesthesia
One of the advantages of...
Analgesia and Pain Management
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Drug Delivery: Parenteral Route
There are three primary parenteral routes: intravenous (IV), intramuscular (IM), and subcutaneous (SC). The IV route introduces the drug directly into the bloodstream, ensuring immediate action. The IM route...