Optical quality of aspheric toric intraocular lenses at different degrees of decentering
Cari Pérez-Vives1, Teresa Ferrer-Blasco, David Madrid-Costa
1GIO, Optics Department, Faculty of Physics, Universidad de Valencia, C/ Dr. Moliner 50., 46100, Valencia, Spain, cari.perez@uv.es.
Purpose:
To analyze the optical quality of aspheric toric intraocular lenses (IOLs) at different degrees of decentering.
Methods:
Wavefront aberrations of Acrysof IQ Toric IOLs (SN6AT3, SN6AT4, and SN6AT5; Alcon Laboratories Inc, Fort Worth, TX, USA) for different powers (15.00, 20.00 and 23.50 diopters [D]) and for different degrees of decentering (diagonal, horizontal, and vertical decentering of 0.3 and 0.6mm) were measured in vitro at 3- and 5-mm pupils. The Zernike coefficients of trefoil, coma, tetrafoil, secondary astigmatism, and spherical aberration were evaluated. The point spread functions (PSFs) and modulation transfer function (MTF) of each IOL evaluated were calculated from the wavefront aberrations.
Results:
Coma aberration increased significantly with IOL decentration. Statistically significant differences were found between centered and all decentered positions in coma aberration (p < 0.05). Although, we only found visible differences between centered and decentered positions in PSF images and MTF curves for horizontal and vertical 0.6 mm of decentering at 3- and 5-mm pupil.
Conclusions:
Despite the coma increment, these values were clinically negligible and have no effect on visual performance, except for vertical and horizontal 0.6 mm of decentering, which could have a negative effect on visual quality.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Degree of Curvature and Radius of Curvature
Properties of Enantiomers and Optical Activity