Measurement of Extracellular Ion Fluxes Using the Ion-selective Self-referencing Microelectrode Technique
Published on: May 3, 2015
A non-linearity criterion applied to the calibration curve method involved with ion-selective electrodes
Tadeusz Michałowski1, Bogusław Pilarski2, Anna M Michałowska-Kaczmarczyk3
1Faculty of Engineering and Chemical Technology, Technical University of Cracow, 31-155 Cracow, Poland.
Abstract:
Some rational functions of the Padé type, y=y(x; n,m), were applied to the calibration curve method (CCM), and compared with a parabolic function. The functions were tested on the results obtained from calibration of ion-selective electrodes: NH4-ISE, Ca-ISE, and F-ISE. A validity of the functions y=y(x; 2,1), y=y(x; 1,1), and y=y(x; 2,0) (parabolic) was compared. A uniform, integral criterion of nonlinearity of calibration curves is suggested. This uniformity is based on normalization of the approximating functions within the frames of a unit area.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
11:30Studying the Activity of Neuropeptides and Other Regulators of the Excretory System in the Adult Mosquito
Published on: August 24, 2021
Related Concept Videos
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Complexometric EDTA Titration Curves
Ion-Exchange Chromatography
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Controlled-Current Coulometry: Overview
Controlled-Current Coulometry: Coulometric Titration
The fundamental requirements for coulometric titrations are (1) 100% efficiency in the reagent-generating electrode reaction and (2) a stoichiometric and preferably rapid reaction between the generated...