The effect of electrode configuration on the unipolar His-bundle electrogram.

A H Kadish1, F Morady, S Rosenheck

  • 1Division of Cardiology, University of Michigan Medical Center, Ann Arbor.

Pacing and Clinical Electrophysiology : PACE
|September 1, 1989
PubMed
Summary

Recording unipolar electrograms from the His bundle position is optimized using an inferior vena cava anode for a better signal-to-noise ratio. This method improves His bundle ablation catheter positioning during electrophysiology studies.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Pain in the perforating type of peptic ulcer.

Journal of the American Medical Association·2010
Same author

Coagulation of the blood in lusteroid tubes; a study of normal persons and patients with arterial or venous thrombosis.

American heart journal·2010
Same author

Coagulation time of the blood in lusteroid tubes; a study of patients receiving dicumarol.

American heart journal·2010
Same author

Pain in the perforating type of peptic ulcer.

The Medical press of Egypt·2010
Same author

Contemporary approach to ablation of paroxysmal atrial fibrillation.

Minerva cardioangiologica·2004
Same author

ACC/AHA clinical competence statement on electrocardiography and ambulatory electrocardiography: A report of the ACC/AHA/ACP-ASIM task force on clinical competence (ACC/AHA Committee to develop a clinical competence statement on electrocardiography and ambulatory electrocardiography) endorsed by the International Society for Holter and noninvasive electrocardiology.

Circulation·2001

Area of Science:

  • Electrophysiology
  • Cardiac Electrophysiology
  • Medical Device Technology

Background:

  • Standardization of unipolar electrogram recording techniques for His bundle ablation is lacking.
  • Current methods for recording His bundle electrograms may not be optimal for catheter positioning.

Purpose of the Study:

  • To evaluate the impact of different anode locations on unipolar electrograms recorded from the His bundle position.
  • To determine the optimal filter settings and anode configuration for recording His bundle electrograms.

Main Methods:

  • Ten patients undergoing electrophysiology studies had unipolar atrial, His bundle, and ventricular electrograms recorded.
  • Five anode locations (chest wall variations, inferior vena cava) and two filter settings (50-500 Hz, 0.05-1000 Hz) were tested.

Related Experiment Videos

  • Signal amplitude, duration, and morphology were analyzed, along with signal-to-noise ratio.
  • Main Results:

    • Anode location did not consistently affect electrogram amplitude, duration, or morphology.
    • The inferior vena cava anode yielded the highest signal-to-noise ratio compared to chest wall anodes.
    • A 50-500 Hz filter setting significantly decreased electrogram amplitudes compared to a 0.05-1000 Hz setting.

    Conclusions:

    • Using an electrode catheter in the inferior vena cava as the anode provides a superior signal-to-noise ratio for unipolar electrograms from the His bundle position.
    • This technique is optimal for recording unipolar electrograms, enhancing accuracy in His bundle ablation procedures.