Related Experiment Video
Updated: Apr 30, 2026

Monitoring the Wall Mechanics During Stent Deployment in a Vessel
Published on: May 8, 2012
Assessment of improved motion-sensitized driven equilibrium (iMSDE) for multi-contrast vessel wall screening
Makoto Obara1, Marc Van Cauteren, Masatoshi Honda
1Philips Electronics Japan Ltd., Healthcare.
Purpose:
We assessed the improved motion-sensitized driven equilibrium (iMSDE) technique for multiple contrast 3-dimensional vessel wall imaging at 3T.
Methods:
Carotid images were obtained using iMSDE combined with turbo field echo (iMSDE-TFE) and conventional double inversion-recovery turbo spin-echo (DIR-TSE) in 5 healthy adult subjects. The tissue signal-to-noise efficiency (SNReff), SNR divided by the square root of scan time, lumen-tissue contrast-to-noise efficiency (CNReff), CNR divided by the square root of scan time, and tissue signal intensity ratio (SIR) were measured in both techniques and compared.
Results:
The iMSDE-TFE had higher SNReff and CNReff (P < 0.01) and strong correlation in SIR (R = 0.92) compared to DIR-TSE.
Conclusion:
The iMSDE-TFE sequence is an efficient technique for vessel wall screening.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT