Use of the Nexfin™ device to detect acute arterial pressure variations during anaesthesia induction
E Weiss1, E Gayat2, V Dumans-Nizard1
1Department of Anaesthesia, Hôpital Foch, University Versailles Saint-Quentin en Yvelines, 92151 Suresnes, France.
Background:
Standard non-invasive arterial pressure (AP) measurements are discontinuous. By providing non-invasive beat-to-beat AP measurements, Nexfin™ might limit duration of intraoperative hypotension and hypertension. We assessed the ability of Nexfin™ to detect AP variations by comparing its trending ability with invasive AP monitoring.
Methods:
Thirty-one subjects undergoing elective surgery under general anaesthesia were included. During induction, simultaneous pairs of AP measurements were collected every 5 s from the Nexfin™ finger sensor and a homolateral radial artery catheter. Magnitude and time lags of AP variations from baseline to nadir and peak were calculated for both methods. Concordance analysis was performed by the Bland-Altman method (for comparison of repeated measures when appropriate).
Results:
Nexfin™ detected 100% of AP changes with the median delays of 0 s (-13 to 7) and 0 s (-5 to 12) for nadir and peak, respectively. Bias [limits of agreement (LOA)] of systolic AP (SAP) variations was -0.5 mm Hg (-31.2 to 30.2) and -9.4 mm Hg (-31.3 to 12.6) from baseline to nadir and from baseline to peak, respectively. For 3479 analysed paired measurements, bias was -3.8 and -8.8 mm Hg for SAP and diastolic AP, with LOA of (-36.0 to 28.5) and (-29.8 to 12.3), respectively.
Conclusions:
Nexfin™ detects AP variations accurately and can be a useful warning device during anaesthesia. However, it is not interchangeable with invasive monitoring, given the large LOA between the two measurements.
Clinical Trial Registration:
NCT01658631.
More Related Videos
11:08Novel Approach for Simultaneous Recording of Renal Sympathetic Nerve Activity and Blood Pressure with Intravenous Infusion in Conscious, Unrestrained Mice.
Published on: February 14, 2018
05:01A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure