Biases in Multicenter Longitudinal PET Standardized Uptake Value Measurements
Robert K Doot1, Larry A Pierce2, Darrin Byrd2
1Department of Radiology, University of Pennsylvania, Philadelphia, PA.
Translational Oncology
|April 29, 2014
Summary
Measurement biases in PET/CT scans can impact therapy assessment. Using standardized calibration sources for PET scanners and dose calibrators helps ensure accurate longitudinal studies and multicenter trials.
Area of Science:
- Nuclear medicine
- Medical imaging
- Radiochemistry
Background:
- Longitudinal positron-emission tomography/computed tomography (PET/CT) studies are crucial for assessing treatment response and conducting multicenter trials.
- Instrumentation variability, including human error, can introduce measurement biases in PET/CT studies.
- Accurate estimation of variability between patient scans is essential for reliable clinical trial outcomes.
Purpose of the Study:
- To investigate measurement biases in longitudinal PET/CT studies caused by instrumentation variability.
- To assess the impact of calibration methodology on standardized uptake value (SUV) errors over time.
- To evaluate the effectiveness of cross-calibration kits for dose calibrators and PET scanners.
Main Methods:
- Utilized National Institute of Standards and Technology-traceable calibration methodology for germanium-68/gallium-68 ((68)Ge/(68)Ga) sources.
- Constructed a cross-calibration kit using a (68)Ge/(68)Ga batch for both dose calibrators and PET scanners.
- Conducted repeat measurements across multiple PET imaging sites using six dose calibrators and six PET/CT scanners.
Main Results:
- Dose calibrator biases ranged from -50% to 9% and were generally stable, except when settings were altered.
- PET scanner biases in activity concentration ranged from -27% to 13% over a median of 174 days.
- Standardized uptake value (SUV) measurement errors ranged from -20% to 47%, with longitudinal biases varying significantly between scanners and over time (-11% to 59%).
Conclusions:
- Measurement biases in SUV are not stable over time and vary between PET scanner sites, impacting longitudinal study reliability.
- Attention to PET scanner calibration is critical for accurate longitudinal assessments.
- Cross-calibration kits for dose calibrators and PET scanners can aid in quality assurance and control for multicenter trials.