Introduction:
There is growing interest in using positron emission tomography (PET) standardized uptake values (SUVs) to assess tumor response to therapy. However, many error sources compromise the ability to detect SUV changes. We explore relationships between these errors and overall SUV variability.
Methods:
We used simulations in a virtual clinical trial framework to study impacts of error sources from scanning and analysis effects on assessment of SUV changes. We varied tumor diameter, scan duration, pretherapy SUV, magnitude of change in SUV, image reconstruction filter, and SUV metric. Poisson noise was added to the raw data before image reconstruction. Variance from global sources of error, e.g., scanner calibration, was incorporated. Two thousand independent noisy sinograms per scenario were generated and reconstructed. We used SUVs to create receiver operating characteristic (ROC) curves to quantify ability to assess response. Integrating area under the ROC curve summarized ability to detect SUV changes.
Results:
Scan duration and image reconstruction method had relatively little impact on ability to measure response. SUVMAX is nearly as effective as SUVMEAN, especially with increased image smoothing and despite size-matched region of interest placement. For an effective variability of 15%, we found the Positron Emission Tomography Response Criteria in Solid Tumors criteria for measuring response (±30%) similar to the European Organization for Research and Treatment of Cancer criteria (±25%).
Conclusions:
For typical PET variance levels, tumor response must be 30% to 40% to be reliably determined using SUVs. PET scan duration and image reconstruction method had relatively little effect.
More Related Videos
10:02Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography FDG-PET/CT
Published on: May 2, 2012
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...