Risks in the implementation and use of smart pumps in a pediatric intensive care unit: application of the failure
Silvia Manrique-Rodríguez1, Amelia C Sánchez-Galindo2, Jesús López-Herce2
1Pharmacy Service,Gregorio Marañón University Hospital.
Insights
Failure Modes and Effects Analysis (FMEA) identified smart infusion pump implementation risks. Improvement measures, including training and monitoring, reduced risk likelihood and improved alarm management in the pediatric intensive care unit.
Area of Science:
- Medical device safety
- Health informatics
- Pediatric intensive care
Background:
- Smart infusion pumps are critical for medication safety in pediatric intensive care units (PICUs).
- Implementation of new technology like smart pumps requires careful risk assessment to ensure optimal use and patient safety.
Purpose of the Study:
- To identify potential risk points during the implementation of smart infusion pumps.
- To prioritize and implement improvement measures for smart pump technology.
Main Methods:
- Failure Modes and Effects Analysis (FMEA) was conducted in a PICU before and 18 months after smart pump implementation.
- A multidisciplinary team assessed risks during drug library creation, clinical use, and data analysis stages.
- Guardrails® CQI v4.1 Event Reporter software was used for data analysis.
Main Results:
- Improvement actions included regular drug library reviews, support document creation, and enhanced alarm distinction.
- These measures reduced the likelihood of risk occurrence and increased detection rates.
- Eighteen months post-implementation, significant risk reduction was observed.
Conclusions:
- FMEA effectively identified and addressed risks associated with smart pump use in the PICU.
- User training, support tools, and continuous monitoring enhanced drug library utility and compliance.
- The study led to improved drug library usefulness, increased user compliance, and a decrease in unnecessary alarms.
Objectives:
The aim of this study was to identify risk points in the different stages of the smart infusion pump implementation process to prioritize improvement measures.
Methods:
Failure modes and effects analysis (FMEA) in the pediatric intensive care unit (PICU) of a General and Teaching Hospital. A multidisciplinary team was comprised of two intensive care pediatricians, two clinical pharmacists and the PICU nurse manager. FMEA was carried out before implementing CareFusion infusion smart pumps and eighteen months after to identify risk points during three different stages of the implementation process: creating a drug library; using the technology during clinical practice and analyzing the data stored using Guardrails® CQI v4.1 Event Reporter software.
Results:
Several actions for improvement were taken. These included carrying out periodical reviews of the drug library, developing support documents, and including a training profile in the system so that alarms set off by real programming errors could be distinguished from those caused by incorrect use of the system. Eighteen months after the implementation, these measures had helped to reduce the likelihood of each risk point occurring and increase the likelihood of their detection.
Conclusions:
Carrying out an FMEA made it possible to detect risk points in the use of smart pumps, take action to improve the tool, and adapt it to the PICU. Providing user training and support tools and continuously monitoring results helped to improve the usefulness of the drug library, increased users' compliance with the drug library, and decreased the number of unnecessary alarms.