Related Experiment Video
Updated: Apr 30, 2026

Subdural Soft Electrocorticography ECoG Array Implantation and Long-Term Cortical Recording in Minipigs
Published on: March 31, 2023
Dural adhesion to porous cranioplastic implant: A potential safety concern
Christina M Sayama1, Mohammad Sorour1, Richard H Schmidt1
1Clinical Neurosciences Center, University of Utah, Salt Lake City, Utah, USA.
Background:
Patient-specific implants are used for cranioplastic skull reconstruction when large bone flaps must be replaced or where there are complex or critical contours, especially near the face. These implants have a low complication rate, with poor fit and postoperative infection being the most common complications. We report here a potentially serious hazard that may arise from the use of porous implants.
Case Description:
A 45-year-old woman sustained severe head trauma in a motor vehicle accident that required urgent surgical intervention. Because of progressive resorption of her native bone flap, she underwent replacement of her native flap with a hard tissue replacement/patient-matched implant cranioplasty. Eight years later, she sustained a traumatic laceration over her vertex that necessitated removal of her cranioplastic implant because of persistent local infection. Intraoperatively, the dural flap was ingrowing and firmly adherent to the inside surface of the porous cranioplasty. After several failed attempts to remove the whole implant piecemeal, we attempted to dissect the dural flap from the brain surface to remove it together with the cranioplastic implant but exposure of the extensive cortical adhesions between the brain surface and the dural flap was compromised by the hard overlying cranioplastic implant. Despite our meticulous attempts to cut off these cortical adhesions, a perisylvian blood vessel was avulsed, resulting in intraparenchymal hemorrhage.
Conclusion:
In this case, dural adhesion and ingrowth to the underside of the cranioplasty implant led to disastrous bleeding when the implant needed to be removed years after initial implantation.
More Related Videos
04:26Author Spotlight: A Reproducible and Efficient Method for Accessing Porcine Brain via Craniectomy
Published on: July 5, 2024
08:42The DREAM Implant: A Lightweight, Modular, and Cost-Effective Implant System for Chronic Electrophysiology in Head-Fixed and Freely Behaving Mice
Published on: July 26, 2024
Related Concept Videos
Increased Intracranial Pressure ll: Pathophysiology
Cranial and Spinal Meninges
Cranial Meninges
These meningeal layers cover the cranium. The dura mater is the outermost layer of cranial meninges. It is a thick and durable membrane of dense...
Cerebral Edema ll: Pathophysiology