Objective:
To develop a capillary electrophoresis system for enantiomeric impurity test of repaglinide.
Methods:
An uncoated fused silica capillary (50 μm×50 cm, with an effective length of 41 cm) was used. The running buffer was composed of 30 mmol/L NaH2PO4 and 5 mg/ml carboxymethyl-β-cyclodextrin(pH 3.5).
Results:
Linear range was 2.00-80.00 μg/ml (correlation coefficient was 0.9993). The average recovery rate was 92.5% to 105.0%.
Conclusion:
The method is simple, accurate and sensitive and it can be used for determination of enantiomeric impurities in repaglinide tablet.
More Related Videos
07:54Analysis of AtHIRD11 Intrinsic Disorder and Binding Towards Metal Ions by Capillary Gel Electrophoresis and Affinity Capillary Electrophoresis
Published on: August 22, 2018
10:33Development of Inhibitors of Protein-protein Interactions through REPLACE: Application to the Design and Development Non-ATP Competitive CDK Inhibitors
Published on: October 26, 2015
Related Concept Videos
Electrophoresis: Overview
There...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary Electrophoresis: Instrumentation
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...